UL94 flame chamber special requirements

The UL94 standard is the standard of a plastic material flame test developed by UL laboratories in the United States. UL94 Tests for Flammability of Plastic Materials for Parts in Devices and Appliances.

This standard has a long history, and ISO, IEC and GB have also adopted this standard.So in the plastics industry, although there are many flame standards, UL94 is still the most basic, the most accepted and used by many customers.

In accordance with the UL94 standard design of the combustion chamber a lot of manufacturers, whether imported or domestic, in fact, and UL laboratory own use, are very different, perhaps this is the UL after long-term use, the experience gained. This is not to say that the existing combustion chamber does not meet the standard provisions, but the actual difference is still very large, so the test results and accuracy are also some differences.

The following brief differences:

Chamber volume: As a standard requirement, a volume of at least 0.5 cubic meters is required and notes that the amount of oxygen available to support combustion is crucial to conducting these tests. For tests that burn for a long time, volumes less than 1 cubic meter may not always provide accurate test results.

Due to production costs, transportation, customer site constraints and other reasons, the general volume of the flame chamber are not more than 1 cubic meters. However, UL laboratory chamber volume, all in the 1.5 cubic meters to 4 cubic meters, showing a great difference.

Sealing: As a standard, the chamber can be completely sealed. However, it does not stress the need to seal or specify the degree of sealing. The general flame chamber, whether the flame door, the operating holes or the top of the exhaust valve, are not sealed, and even for convenience, many of the operations are open the door. The UL laboratory for the seal, is exceptionally stringent, generally two ways to verify, one candle method, the second is the smoke bomb method. It should be stressed that, generally, there is little concern about the sealing of exhaust valves, but this is a factor that greatly affects the test results. Because the laboratory exhaust pipe is always connected to the chamber, if the tightness is not good, pressure changes in the pipe (such as the fan always exhausting, outdoor air disturbance, etc.) will interfere with the normal test in the chamber.

Gas flow meter: although the standard does not require a specific structure, but basically because of cost considerations, are the use of rotameters. The shortcomings of this flowmeter is obvious, readings are relative, you must use methane gas to calibration in order to get the exact scale (which is the majority of calibration institutions can not be achieved), the actual flow with the temperature needs to be amended. Even with the use of rotor flowmeters, UL has its own brand of long-term use recognized internally. However, UL is no longer using this type of flowmeter, instead of mass flowmeter which could be set to the type of flow parameters of the quality of gas. According to sources, UL94 standards in the future may be amended to provide for the use of such mass flowmeter.

Bunsen burner: The standard does not require the use of the brand, but provides must comply with ASTM D5025. UL also has its own long-term use and recognition of the brand. Because of the structure of the burner is more complicated, high precision parts, not the general manufacturers to produce. Random use of non-standard burner, will result in extreme instability of the flame, making the test results deviate.

Thermocouple: Thermocouples are important instruments used to verify flame power. The quality of the thermocouple itself as well as the copper slug attached to the thermocouple material, size, weight, fixing methods and other requirements are extremely strict. The power of the flame is extremely important to the test result. When the test conditions change and the replacement of methane gas,  the flowmeter, burner, back pressure and other parameters need to be adjusted.