Rh-4019 programmable constant temperature and humidity test chamber
Applicable Scope
Constant temperature and humidity test chamber can program is suitable for the aerospace, electronic electrician, ship railway, automobile motorcycle, quality measurement and scientific research units, institutions of higher learning, enterprises and institutions, all kinds of electronic elements and other products of spare parts and materials in high temperature, low temperature, constant temperature and stored under hot and humid environment and the adaptability when using test, test their various performance indicators, is the product test simulation environment necessary testing equipment.
Technical Characteristics
1. High quality appearance, the body adopts arc shape, the surface is treated by the stripe of the mist surface, and the plane has no reaction handle, easy to operate, safe and reliable.
2. Rectangular laminated glass viewing window, which can be used for observation of test products in the test. The window is equipped with anti-perspirant electric heater device to prevent water vapor condensation droplets, and high-brightness PL fluorescent lamp to keep the lighting inside the box.
3. The door is sealed with double layers to prevent air tightness and effectively prevent internal temperature leakage.
4. With an external water supply system, it is convenient to replenish the water supply from the humidifying bucket, and can be recycled automatically.
5. The compressor circulation system adopts the French "taikang" brand, which can effectively remove the lubricating oil between the condensing pipe and the capillary pipe, and the whole series adopts the environmental protection refrigerant (R23, R404, R507).
6. The controller adopts imported LCD display screen, which can display the measured value, set value and time at the same time.
7. The controller has multiple program editing and temperature and humidity can be quickly (OUICK) or slope (SLOP) control.
8. The built-in movable pulley is easy to move and put and has a strong positioning screw fixed position.
Meet Standards
GB/T 10586-2006 technical conditions for wet heat test chamber
GB/T 10589-2008 low-temperature test chamber technical conditions
Technical conditions for high and low temperature test chamber
Technical conditions for high temperature test chamber
Environmental tests for electrical and electronic products -- part 2: test methods -- test A at low temperature
IEC 60068-2-2007 environmental tests for electrical and electronic products -- part 2: test methods -- test B at elevated temperatures
Environmental tests for electronic and electronic products -- part 2: test methods test Cab: constant wet heat test
Environmental testing of electrical and electronic products -- part 2: test methods test Db alternating wet heat (12h+12h cycle)
Requirements for performance verification standards for environmental temperature and humidity test Chambers for electrical and electronic products
Requirements for methods for verification of basic parameters of combined temperature/humidity cycle test equipment for electronic products
IEC 60068-2-14-2009 environmental tests for electrical and electronic products -- part 2: test N: temperature change
Guide for environmental testing of electrical and electronic products at high temperature and low temperature
GJB 150.3 high temperature test
GJB 150.4 low temperature test
GJB 150.9 wet heat test
Measurement requirements:
Temperature can be controlled at 20 ℃ ~ 60 ℃, humidity can achieve 95% plus or minus 3%. The expanded uncertainty of the temperature measure system (k = 2) not more than 0.4 ℃; The thermal time constant of the sensor :20s~40s, the extended uncertainty of the humidity measurement system (k=2) is not greater than 1/3 of the allowable deviation of the humidity under test.
Performance Performance
System
|
Balanced temperature & humidity control system
|
Temperature range
|
-40~+150℃
|
Humidity range
|
20~98%RH
|
Temp.fluctuation
|
≤1℃
|
Temp. uniformity
|
≤2℃
|
Temp. variation
|
≤1℃
|
Humid. variation
|
+2/-3.0%RH
|
Heat up time
|
-40→+150℃ 约40min
|
Pull down time
|
+25→-40℃ 约60min
|
Components
|
Interior material
|
SUS304 Stainless steel plat SUS304
|
Exterior material
|
Stainless steel plat
|
Heat insulating material
|
Rigid polyurethane foam+glass wool
|
Observation window
|
Glass incorporating heat generator
|
Cable hole
|
Φ50 Inside diameter Φ50
|
Heater
|
Iron-chrome wire heater
|
Humidifier
|
Copper sheathed heater
|
Circulating fan
|
Multi-wing centrifugal blower
|
Motor
|
Condenser type
|
Caster wheel
|
Lockable caster wheel
|
Sensor
|
PT100Ω
|
Controller
|
Controller
|
Touch screen type LCD Monitor
|
Setting method
|
Touch screen monitor
|
Control method
|
PID控制PID control by microprocessor
|
Programs
|
120组程序1200段,全部循环99次 120 programs 1200 steps, all repeat 99 cycle
|
Operation function
|
Fixed operation, Programmable operation
|
Refrigeration system
|
Refrigerating mode
|
Mechanical single-stage refrigerator system
|
Air-cooled condenser
|
Refrigerator
|
Hermetically sealed rotary compressor
|
Refrigerant
|
Non-chlorofluorocarbon Coolant R404A
|
Condenser
|
Finned heat exchanger
|
Evaporator
|
Finned tube Evaporator
|
Water supply
|
Water supply method
|
Water pump out system
|
口Add water entrance
|
口Water entrance before the chamber
|
Storage water tank
|
Stainless steel water tank
|
Safe Device
|
Independence overstep. protection/over current breaker/ Refrigerator overload relay, refrigerator overpressure and overheat protection.
|
Standard
|
Interior dimensions
|
400×500×500mm(D×W×H)
|
Exterior dimensions
|
850×1000×1430mm(D×W×H)
|
Internal volume(L)
|
100L
|
Power supply
|
AC380V/50Hz
|
Power
|
约4.5KW(-40~+150℃)
|
Optional Accessories
|
Additional shelf plate/Communication interface RS232C.
|
|